A special phantom for Quality Assurance of Digital Subtraction Angiography Equipment (DSA) according to IEC 61223-3-3 and DIN 6868/Parts 8, (2007) and 54. The phantom contains a movable insert with aluminium strips to simulate blood vessels. This insert can be moved via pneumatic system from a safe distance during exposure.
It can be used to measure:
- Artifacts
- Dynamic Range
- Logarithmic Errors
- Contrast Media Sensitivity
Features and Benefits
- Total PMMA thickness in the test area: 57mm
- Dynamic 7 steps wedge of copper with thicknesses from 0.2 mm to 1.4 mm arranged perpendicular to the longitudinal direction of the insert
- Additional step from 1.4mm to 0.2mm for compensation test
- Additional step from 0.0mm to 0.2mm
- Movable PMMA insert with 4 aluminum strips (0.05, 0.1, 0.2, 0.4mm), purity 99.5%, simulating different vascular densities
- Pneumatic 6m hose with a valve allowing for remote movement of the insert in both directions
Technical Specifications
- Dimensions: 8′ x 7′ x 2′
Warranty Information
- 5-Year Warranty